Reportable based on device analysis completed on 11jun2021.It was reported that the balloon did not inflate at all.The target lesion was located in a mildly tortuous artery.A 4.00mm/1.5cm/140cm small peripheral cutting balloon was selected for use.During the procedure, after the balloon was delivered to the lesion, inflation was tried to perform.However, it was noted that the balloon did not inflate at all.The procedure was completed with a different device.No patient complications were reported.However, device analysis revealed that there was a balloon pinhole.
|