The instrument was manufactured in (b)(6) 2016.Upon examination, it can be seen that the distal tip was broken and has no discoloration on a broken surface; there is some rust on the stopcock.Multiple mechanical damages observed in the metal tubing such as bent, dented.The root cause of breakage is most likely due to mechanical overload i.E.Use error.No indication for a material or manufacturing related issue was found during the investigation.Ifu includes warning of risk or injury: overloading the instrument by exerting too much force may cause the medical device to break, bend and malfunction, and consequently injure the patient.
|