Model Number 4542
Device Problems Signal Artifact/Noise (1036); Failure to Capture (1081); Fracture (1260); High impedance (1291); High Capture Threshold (3266)
Patient Problem Heart Failure/Congestive Heart Failure (4446)
Event Date 04/07/2021
Event Type  Injury  
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.
 
Event Description
It was reported that this patient was admitted into the hospital with heart failure symptoms.During a normal battery generator change out procedure, this left ventricular (lv) lead exhibited noise, loss of capture, high threshold and high out of range pacing impedance, measuring greater than 3000 ohms.Lead fracture was suspected.The physician decided not to proceed with a new lv lead replacement but instead programmed the new device to vvir, right ventricular only.No additional adverse patient effects were reported.This lv lead remains implanted and left in situ.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12111872
MDR Text Key259983926
Report Number2124215-2021-13512
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526410901
UDI-Public00802526410901
Combination Product (y/n)Y
Reporter Country CodeAS
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 07/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/07/2015
Device Model Number4542
Device Catalogue Number4542
Device Lot Number158501
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/07/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/07/2013
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
-
-