MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN CRT-D; IMPLANTABLE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN CRT-D; IMPLANTABLE DEVICE Back to Search Results
Model Number G140
Device Problem Impedance Problem (2950)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/07/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the cardiac resynchronization therapy defibrillator was explanted and replaced due to a shock impedance issue.No additional adverse patient effects were reported.
 
Event Description
It was reported that the cardiac resynchronization therapy defibrillator was explanted and replaced due to a shock impedance issue.No additional adverse patient effects were reported.
 
Manufacturer Narrative
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN CRT-D
Type of Device
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12112225
MDR Text Key260024618
Report Number2124215-2021-17404
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534539
UDI-Public00802526534539
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/02/2018
Device Model NumberG140
Device Catalogue NumberG140
Device Lot Number481938
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/14/2021
Initial Date FDA Received07/02/2021
Supplement Dates Manufacturer Received05/11/2022
Supplement Dates FDA Received06/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age73 YR
Patient SexMale
-
-