Model Number G447
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be submitted.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was electively replaced due to head connection issues.There is no further information of the issue at this time.The device was received for analysis.No additional adverse effects reported.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was electively replaced due to head connection issues.No additional adverse effects reported.
 
Manufacturer Narrative
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
IMPLANTABLE DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key12114145
MDR Text Key260269204
Report Number2124215-2021-18120
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589539
UDI-Public00802526589539
Combination Product (y/n)N
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 09/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/12/2020
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number187034
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2021
Initial Date Manufacturer Received 06/02/2021
Initial Date FDA Received07/03/2021
Supplement Dates Manufacturer Received09/10/2021
Supplement Dates FDA Received09/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-