Catalog Number 8607000 |
Device Problems
Gas Output Problem (1266); Failure to Deliver (2338); Pressure Problem (3012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/03/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still on-going.The results will be provided with a follow-up report.
|
|
Event Description
|
It was reported that during a case the unit displayed a ventilator failure.There was no patient injury.
|
|
Event Description
|
It was reported that during a case the unit displayed a ventilator failure.There was no patient injury.
|
|
Manufacturer Narrative
|
The dispatched fse tested the device and was able to duplicate the reported observation; the workstation was unable to build up the auxiliary vacuum pressure.A new vacuum pump has been installed as well as all parts that are included in the 2 years preventive maintenance kit (upper and lower piston diaphragms, seals and o-rings).The device passed all tests and could be returned to use.The auxiliary vacuum pressure is needed to operate the valves in the cosy and to keep the ventilator diaphragms in place during piston movement to avoid wrinkling.If the pump cannot build-up the necessary pressure the device is designed to shut down automatic ventilation to prevent from serious damages to the ventilator unit resulting from a jammed piston diaphragm and to avoid potentially hazardous output because the valves for ventilation control cannot be actuated properly.The shut-down is accompanied by a corresponding vent fail alarm.All parts that could be put in causal connection to a loss of vacuum pressure have been replaced.It was not exactly determined which part indeed caused the leakage in this specific pneumatic circuit but lack of maintenance cannot be ruled out as a contributing factor.
|
|
Search Alerts/Recalls
|