• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIGMA HP UNI FEM FIN BLK SZ 5; KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIGMA HP UNI FEM FIN BLK SZ 5; KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES Back to Search Results
Catalog Number 202471500
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Bone Fracture(s) (1870)
Event Date 06/16/2021
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the male right medial uni had a medial facet fracture of the trochlea and a screw was required to contain.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Product and lot number of the reported devices (anterior chamfer osteotome and femoral prep block) 2024-71-500 and 2024-85-000.Was there any alleged deficiency to the reported devices? if yes, please provide details.No.Please verify if the instrument used during surgery? yes femoral finishing block and anterior chisel.If yes, was there surgical delay? yes.What was the duration of the delay? 10 minutes whilst they inserted a screw.The hospital will not release the instrumentation until we can provide replacements so that scheduling and theatre lists are not impacted.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device was received for examination, therefore the reported event could not be confirmed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI FEM FIN BLK SZ 5
Type of Device
KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12121547
MDR Text Key260041581
Report Number1818910-2021-14324
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 07/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number202471500
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/16/2021
Initial Date FDA Received07/06/2021
Supplement Dates Manufacturer Received07/15/2021
08/30/2021
Supplement Dates FDA Received08/01/2021
08/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-