MAUDE Adverse Event Report: MEDTRONIC INC STEALTHSTATION FLEXENT; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC INC STEALTHSTATION FLEXENT; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9736242
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2021
Event Type  malfunction  
Manufacturer Narrative
Onsite functional and visual examination was performed by a manufacturer representative.The system passed a system checkout and was determined to be operational.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation device being used for a functional endoscopic sinus surgery (fess).It was reported that the site was demonstrating a trace pattern to residents.When they went to restart the trace pattern to demonstrate the pattern again the system screen went black and they received an internal error 64.They clicked on the prompt and the system said a reboot was needed.The site clicked ok and the system reboot.After the reboot there were no issues and they continued with the case.There was no reported delay and no reported impact to patient outcome.
 
Manufacturer Narrative
Continuation of d11: section d information references the main component of the system.Other relevant device(s) are: product id: 9736226 software, serial/lot #: version 1.3.2-20, ubd:n/a, udi#: h3/h6) the system logs from this event were provided for software analysis.The logs were reviewed and the reported issue was related to a software issue.This issue has been documented in a navigation software anomaly tracking database.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION FLEXENT
Type of Device
EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC INC
826 coal creek cir #100
louisville CO 80027
MDR Report Key12129494
MDR Text Key260299707
Report Number1723170-2021-01786
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
PMA/PMN Number
K200723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 09/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9736242
Device Catalogue Number9736242
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/16/2021
Initial Date FDA Received07/07/2021
Supplement Dates Manufacturer Received07/13/2021
09/29/2021
Supplement Dates FDA Received07/13/2021
09/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SEE H10...
Patient Age28 YR
Patient Weight125
-
-