Model Number 0998-00-3023-53 |
Device Problem
Failure to Run on Battery (1466)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device: (10/3233) the getinge field service engineer (fse) who encountered the issue replaced the power supply to address the reported issue.The fse performed all functional and safety tests were passed to meet factory specifications, and the iabp was returned to the customer and cleared for clinical service.Upon completion of our investigation, a supplemental report will be submitted.(b)(6).
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse) the getinge loaner cs300 intra-aortic balloon pump would not operate on battery power.There was no patient involvement and no adverse event was reported.
|
|
Manufacturer Narrative
|
Analysis of production: (3331/213) the device history record review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period jul 2019 through jun 2021 was reviewed.There were no triggers identified for the review period.
|
|
Event Description
|
N/a.
|
|