Model Number 71992-01 |
Device Problems
Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
|
Patient Problems
Hypoglycemia (1912); Loss of consciousness (2418); Convulsion/Seizure (4406); Cramp(s) /Muscle Spasm(s) (4521)
|
Event Date 06/22/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the freestyle libre sensor.Customer received a "scan again in 10" message for more than two hours and was unable to obtain readings.As a result, customer experienced cramping, a seizure, and loss of consciousness and was unable to self-treat, requiring treatment of baqsimi 3mg nasal powder.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
No product has been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kits were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the freestyle libre sensor.Customer received a "scan again in 10" message for more than two hours and was unable to obtain readings.As a result, customer experienced cramping, a seizure, and loss of consciousness and was unable to self-treat, requiring treatment of baqsimi 3mg nasal powder.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.The sensor plug is properly seated and no physical damage was observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.Sim vivo testing (simulation of the electrical signal produced by the sensor tail) was performed and all results were within specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the freestyle libre sensor.Customer received a "scan again in 10" message for more than two hours and was unable to obtain readings.As a result, customer experienced cramping, a seizure, and loss of consciousness and was unable to self-treat, requiring treatment of baqsimi 3mg nasal powder.There was no report of death or permanent impairment associated with this event.
|
|