Catalog Number 0250070520 |
Device Problems
Contamination /Decontamination Problem (2895); Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.
|
|
Event Description
|
It was reported that foreign material was found inside the sterile packaging.
|
|
Manufacturer Narrative
|
Alleged failure: there is a piece of black debris (potentially a bug) in the sterile packaging.There is also a sticky blue substance on the tubing the failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be the tyvek cover was dirty or the packagers hand had contaminants during packaging.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
|
Event Description
|
It was reported that foreign material was found inside the sterile packaging.
|
|