Model Number 1035-41-000 |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the 41mm bipolar head poly insert was loose and had too much toggle with the outer metal shell.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # : (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : a device history record (mre) review, was not possible because the required lot code was not provided.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).Investigation summary: device associated with this report was received for examination.Complain is not confirmed.A dimensional inspection was performed.And all dimensions are confirming, in accordance to specification.Depuy synthes considers, the investigation closed at this time.Should additional information be received, the information will be reviewed.And the investigation may be re-opened as necessary.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device part: 103541000, lot#: ja1018.And no non-conformances/manufacturing irregularities were identified.Device history review: a manufacturing record evaluation (nc search) was performed for the finished device [lot/serial/batch] number.And no non-conformances/ manufacturing irregularities were identified.
|
|
Search Alerts/Recalls
|