• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number RA002-4545SL
Device Problems Corroded (1131); Retraction Problem (1536)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to the (b)(4) implant retrieval centre (lirc); no product was returned to the manufacturer for investigation.A root cause was unable to be determined with the information provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event Description
It was reported that a rod was explanted on an unknown date for an unknown reason.Subsequently, upon review of investigation findings from the (b)(4) implant retrieval centre (lirc), it was identified that the rod exhibited signs of corrosion, debris present within the housing tube, and was unable to retract.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. suite 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. suite 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr. suite 100
aliso viejo, CA 92656
8583448112
MDR Report Key12139818
MDR Text Key260800276
Report Number3006179046-2021-00385
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00856719002183
UDI-Public856719002183
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 07/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRA002-4545SL
Device Lot NumberA140609-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/25/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/09/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-