• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY MEMORY II DOUBLE LUMEN EXTRACTION BASKET; LQR DISLODGER, STONE, BILIARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY MEMORY II DOUBLE LUMEN EXTRACTION BASKET; LQR DISLODGER, STONE, BILIARY Back to Search Results
Catalog Number MB-35-2X4-8
Device Problems Difficult to Fold, Unfold or Collapse (1254); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/24/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report.The basket was returned fully retracted into the catheter.When the handle was manipulated the basket would not move.The handle was disassembled and it was noted that the drive wire had disconnected from the wire.There was nesting noted in the purple hub.For further evaluation of the drive wire cable and basket, the catheter was cut to push the drive wire cable out of the sheath.The basket was fully formed and intact, and evidence of solder was present on the handle cannula at the joint.No other anomalies were detected with the devices.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the report of unable to open the basket was due to a separation of the device in the handle.The cause of the separation is unknown.The instructions for use indicates: "advance device through channel, in short increments, until basket sheath exits endoscope." the instructions for use also states: "confirm desired position of basket sheath relative to target.Advance basket out of sheath.Caution: pulling on sheath while advancing or retracting basket may damage device, rendering it inoperable." basket deployment difficulties and buckling/breaking of the drive wire can occur if the device experiences excessive pressure.Resistance in basket movement and bends in the catheter can occur if the elevator of the endoscope is used to deflect the device at a sharp angle.Prior to distribution, all memory ii double lumen extraction baskets are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook memory ii double lumen extraction basket.The physician tried to remove the stones in the common bile duct.The assistant nurse once captured the stone it worked well but after several times of handling, the handle didn't move at all.The nurse removed the basket out of the endoscope channel and used another basket.There was no reportable information at this time.Our evaluation of the returned device on 15-jun -2021 determined that the drive wire had disconnected from the wire with nesting of the wire inside the purple hub [subject of the report].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEMORY II DOUBLE LUMEN EXTRACTION BASKET
Type of Device
LQR DISLODGER, STONE, BILIARY
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
kaylin davis
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key12141844
MDR Text Key261940472
Report Number1037905-2021-00303
Device Sequence Number1
Product Code LQR
UDI-Device Identifier10827002251493
UDI-Public(01)10827002251493(17)230909(10)W4383638
Combination Product (y/n)N
PMA/PMN Number
K171969
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/09/2023
Device Catalogue NumberMB-35-2X4-8
Device Lot NumberW4383638
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2021
Initial Date Manufacturer Received 06/14/2021
Initial Date FDA Received07/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-