UTAH MEDICAL PRODUCTS, INC. UMBILI-CATH; CATHETER, UMBILICAL ARTERY Back to Search Results
Model Number SINGLE-LUMEN SILICONE
Device Problems Crack (1135); Gas/Air Leak (2946)
Patient Problem Insufficient Information (4580)
Event Date 04/02/2021
Event Type  malfunction  
Event Description
Uac (umbilical artery catheter) stopcock found cracked after noting air in the line.This is a disposable line with a crack causing air to get into umbilical line.Air was seen and removed the staff.No air got through the line to the baby.Stopcock changed and cracked one removed.Checked par carts and central stock, did not find any affected lot# 1202974.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UMBILI-CATH
Type of Device
CATHETER, UMBILICAL ARTERY
Manufacturer (Section D)
UTAH MEDICAL PRODUCTS, INC.
7043 south 300 west
midvale UT 84047
MDR Report Key12142246
MDR Text Key260725159
Report Number12142246
Device Sequence Number1
Product Code FOS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberSINGLE-LUMEN SILICONE
Device Catalogue Number4173505
Device Lot Number1202974
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA06/24/2021
Event Location Hospital
Date Report to Manufacturer07/09/2021
Type of Device Usage Unknown
Patient Sequence Number1
-
-