|
Model Number 9553 |
Device Problem
Positioning Failure (1158)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/28/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The promus premier ous mr 32 x 2.25mm stent delivery system was returned for analysis.A visual examination of the stent found evidence of distal stent damage with stent struts bunched proximally.The undamaged crimped stent outer diameter was measured and the result is within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip found distal tip damage.A visual and tactile examination of the hypotube found no issues.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.The device was soaked over-night at 37 degrees to facilitate functional testing.A recommended 0.014 guidewire was loaded successfully before the functional test.The balloon was inflated to rated burst pressure without issue, maintained pressure and the stent expanded successfully despite the stent damage.The device was deflated successfully in 11 seconds which is within deflation time specification.The inflation device was verified before and after use.No other issues were identified during the product analysis.
|
|
Event Description
|
Reportable based on device analysis completed on 24-jun-2021.It was reported that deployment failure occurred.The patient presented with myocardial infarction.The stenosed, 32 x 2.40mm target lesion was located in the mildly calcified and moderately tortuous distal left anterior descending artery.A 32 x 2.25mm promus premier drug-eluting stent was advanced but the stent was unable to expand.The procedure was completed with another of the same device.No patient complications were reported and the patient status was stable.However, returned device analysis revealed stent damage.
|
|
Search Alerts/Recalls
|
|
|