MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA¿ QUAD XT CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTBA1QQ |
Device Problems
Mechanical Problem (1384); Misconnection (1399)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/08/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant products: 5076-45 lead implanted (b)(6) 2008.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) exhibited a grommet or setscrew problem as the lead could not be removed from the device header.A bone cutting device was used to open the header and remove the lead which was functioning normally.The crt-d was explanted and replaced.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.The returned device indicated shorting/low impedance in the battery.B attery analysis revealed that the decrease in the voltage of battery was likely due to an electrochemical short due to the growth of ag dendrites.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.The returned device indicated shorting/low impedance in the battery.Battery analysis revealed that the decrease in the voltage of battery was likely due to an electrochemical short due to the growth of ag dendrites.The decrease in the voltage of battery was likely due to cathode particle induced short.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|