Model Number SL-2010M2096 |
Device Problems
Leak/Splash (1354); Detachment of Device or Device Component (2907)
|
Patient Problem
Hemorrhage/Bleeding (1888)
|
Event Date 06/30/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
|
Event Description
|
As reported by user facility: while attempting to adjust a kink in the venous tubing leading to the dialyzer, the tubing came apart at the blue dialyzer connector.Blood could not be returned to the patient.A new setup was completed and the treatment resumed.200ml of blood loss for patient.
|
|
Manufacturer Narrative
|
This report has been identified as b.Braun medical inc.Internal report number (b)(4).All information associated with this event was submitted to the bloodline manufacturer.According to the manufacturer investigation, three (3) blood tubing sets were returned for product evaluation.The complaint was confirmed with two of the sets with visual inspection finding the blue dialyzer connector was disconnected from the tubing.No issues were observed with the third tubing set.Photographs of the two (2) disconnected sets were taken and provided to manufacturing for review.Manufacturing was unable to determine a root cause for this issue based on the available information.A review of the device history records for sl-2010m2096 from lot 10354003 was performed and indicated that there were no quality issues during the manufacturing process of this lot related to the reported issue.If additional pertinent information becomes available a follow-up report will be filed.
|
|
Search Alerts/Recalls
|