MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME Back to Search Results
Model Number N/A
Device Problems Failure to Cut (2587); Patient Device Interaction Problem (4001)
Patient Problems Abrasion (1689); Scar Tissue (2060)
Event Date 06/16/2021
Event Type  Injury  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Event Description
It was reported that during surgery the patient experienced a bad cut as the unit was not cutting properly.Patient had 2 separate pieces of skin with an absence in the area of the ¿racing stripe." no other adverse event was reported as a result of this event.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Event Description
Grafts required additional manipulation by surgical team to utilize harvested graft.Patient did not require additional harvesting sites and did not experience extensive injury.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record for pn 00882100100 determined the rpm¿s were in specification, but erratic and the calibration was out.The adjustment cams and semi-circle and vespel bearings were replaced to aid in calibration.The spring seal was stretched to help the erratic rpms.Devices are used for treatment.A definitive root cause cannot be determined.The event cannot be confirmed.
 
Event Description
There is no additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key12159247
MDR Text Key261257049
Report Number0001526350-2021-00766
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number63258035
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/17/2021
Initial Date FDA Received07/13/2021
Supplement Dates Manufacturer Received07/28/2021
11/15/2021
Supplement Dates FDA Received07/30/2021
11/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/16/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-