Model Number N/A |
Device Problems
Failure to Cut (2587); Patient Device Interaction Problem (4001)
|
Patient Problems
Abrasion (1689); Scar Tissue (2060)
|
Event Date 06/16/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that during surgery the patient experienced a bad cut as the unit was not cutting properly.Patient had 2 separate pieces of skin with an absence in the area of the ¿racing stripe." no other adverse event was reported as a result of this event.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
Grafts required additional manipulation by surgical team to utilize harvested graft.Patient did not require additional harvesting sites and did not experience extensive injury.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record for pn 00882100100 determined the rpm¿s were in specification, but erratic and the calibration was out.The adjustment cams and semi-circle and vespel bearings were replaced to aid in calibration.The spring seal was stretched to help the erratic rpms.Devices are used for treatment.A definitive root cause cannot be determined.The event cannot be confirmed.
|
|
Event Description
|
There is no additional information.
|
|