Model Number N/A
Device Problems Failure to Cut (2587); Patient Device Interaction Problem (4001)
Patient Problems Abrasion (1689); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Event Description
It was reported that during a procedure blade created an incomplete graft (1 piece of skin with a thinned racing stripe ¿ [10 10 10 10 10 6 10 10] ).No other adverse events were reported as it relates to this event.
 
Event Description
It was reported that the graft was less than optimal.This resulted in the patient having harvest sites that might not heal without additional wound management.The graft also required additional manipulation by the surgical team to utilize the harvested graft.Fortunately, the patient did not require additional harvesting site and did not experience extensive injury.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, g3, g6, h2, h3, h4, h6, h10.Visual examination of the provided pictures identified that the blade and dermatome were used, and the graft taken was thinner in one spot.Devices are used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Event Description
There is no additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key12159646
MDR Text Key261289974
Report Number0001526350-2021-00793
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number64063483
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-