The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information and records received, the reported event of unable to interrogate via ble was confirmed.The logs were reviewed and determined there was a message integrity check (mic) error which resulted in connection errors, which ultimately led to a cyber lockout.The cause of the mic error is underdetermined.Based on the result of these findings, abbott is performing further investigation.
|