• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE MEDICAL SYSTEMS, LLC SIGNA PIONEER; NUCLEAR MAGNETIC RESONANCE IMAGING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS, LLC SIGNA PIONEER; NUCLEAR MAGNETIC RESONANCE IMAGING Back to Search Results
Device Problems Thermal Decomposition of Device (1071); Use of Device Problem (1670)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2021
Event Type  malfunction  
Manufacturer Narrative
Unique identifier: (b)(4).There are no additional device identification numbers.Ge healthcare's investigation is ongoing.A follow up report will be submitted once the investigation has been completed.Device evaluation anticipated, but not yet begun.
 
Event Description
It was reported that while preparing the room for an mr exam and prior to patient entry, the technologist noticed a small section of the table appeared to be burned.
 
Manufacturer Narrative
H3: the investigation by ge healthcare (gehc) has been completed.The mr system was operating within specifications and all safety mitigating devices were functional when checked by the gehc field engineer.The third-party vendor found no deficiencies in the gems flex coil.The root cause of the incident was misuse by setting up the coil in a way that does not follow cable routing safety warnings as described in the operator documentation.The operator manual and warnings provide that crossing or looping of cables may result in burns to the patient.The mr operator has the final responsibility for the use and placement of the coil set-up and preparation of the patient, prior to starting the mr exam procedure.No further actions are planned by gehc.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGNA PIONEER
Type of Device
NUCLEAR MAGNETIC RESONANCE IMAGING
Manufacturer (Section D)
GE MEDICAL SYSTEMS, LLC
3200 n grandview blvd.
waukesha, WI 53188
MDR Report Key12162397
MDR Text Key263413840
Report Number2183553-2021-00007
Device Sequence Number1
Product Code LNH
Combination Product (y/n)N
PMA/PMN Number
K143345
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-