Model Number N/A |
Device Problems
Power Problem (3010); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/09/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that it sounds strange according to or personnel, no patient was affected.Medical technical department have taken a look and wanted to send it in for maintenance because it's slow when starting the machine.The event occurred when checking the machine prior to surgery.No adverse event was reported as a result of this malfunction.
|
|
Event Description
|
There is no additional information.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that unit was partially seized and made a rattling noise.The bearings and seal were replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|
|
Search Alerts/Recalls
|