• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.; LIGATION / VESSEL CLIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.; LIGATION / VESSEL CLIPS Back to Search Results
Model Number PL574T
Device Problem Unintended Ejection (1234)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/19/2021
Event Type  malfunction  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event Description
It was reported to aesculap ag that a challenger ti-p sm-ligat.Clips 12 cartr.(part # pl574t) was used during a laparoscopic hepatectomy procedure performed on (b)(6) 2021.According to the complainant, during the procedure, the cartridge detached.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the event.Although requested, additional information has not been made available.The malfunction is filed under aag reference (b)(4).
 
Manufacturer Narrative
Investigation results: visual investigation: the investigation was carried out visually and microscopically with the digital microscope and the digital-camera.The investigator made a visual inspection of the products.Here they detected a broken off nose, wrong positioned clips, bent latches and deformations on the slider sheet.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Conclusion and measures / preventive measures: according to the quality standard and dhr files a material defect and a production error can be excluded.There are no hints of a pre-damage or similar.Investigations lead to the assumption that the cartridge is not engaged completely, there is an impairment of product functionality.This could led to deformed latches of the slider, to wrong positioned clips and a clip jam.A clip jam could also lead to a detached cartridge.Furthermore a too fast application of the cartridge can also led the clip jam and deform the slider sheet.Based upon our historically grown product experience and due to different simulation regarding a cartridge which is engaged not completely, this leads to the described errors.If the shaft / applier no longer conforms to the specification, this can also result in an incorrect engaged cartridge.According to the investigation results there is no capa request necessary.
 
Event Description
Update: this case is associated wit following mdr number 9610612-2021-00500 (400518301 pl574t), 9610612-2021-00539 (400518302 pl574t), 9610612-2021-00501 (400518303 pl572t).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.
Type of Device
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key12166069
MDR Text Key261487904
Report Number9610612-2021-00500
Device Sequence Number1
Product Code FZP
UDI-Device Identifier04046963356057
UDI-Public4046963356057
Combination Product (y/n)N
PMA/PMN Number
K081031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL574T
Device Catalogue NumberPL574T
Device Lot Number52622401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2021
Initial Date Manufacturer Received 06/21/2021
Initial Date FDA Received07/14/2021
Supplement Dates Manufacturer Received07/19/2021
Supplement Dates FDA Received08/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-