Catalog Number MZ1000
Device Problem Blocked Connection (2888)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2021
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: unknown.Device manufacture date: unknown.Investigation summary: one mz1000 sample was received without packaging for investigation.No connecting products were returned to assist the investigation.A visual inspection of the returned sample did not identify any product defects or manufacturing defects which could have contributed to the customer's experience.Functional testing was performed by connecting a retained 50ml bd plastipak syringe and a three way tap from stock to the maxzero component; in each instance a secure connection was observed and no disconnections were identified.The details of this feedback were forwarded to the manufacturing site for investigation.In this instance, a definitive root cause could not be identified as testing of the returned samples did not identify any product defects that could have contributed to the customer¿s experience.The lot number was not available and therefore it is not possible to perform a review of the production documentation for this particular product.A review of the customer feedback database indicates that complaints of this nature are rare and there is currently no trend for issues of this nature against the mz1000 product.
 
Event Description
It was reported that maxzero needleless connector was loose.The following information was provided by the initial reporter: this is a report about a connection issue of mz1000.According to the customer's report, the connection of the connector to the iv line is loose.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXZERO NEEDLELESS CONNECTOR
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12170476
MDR Text Key261735967
Report Number9616066-2021-51521
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberMZ1000
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/15/2021
Initial Date FDA Received07/14/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-