MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SYRINGE IRRIGATION BULK / CLEAN, 50ML; IRRIGATION SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SYRINGE IRRIGATION BULK / CLEAN, 50ML; IRRIGATION SYRINGE Back to Search Results
Model Number 0935280
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that there was hair found in the syringe.
 
Manufacturer Narrative
The reported event was confirmed cause unknown.1 sample was confirmed to exhibit the reported failure.The device had not met specifications.The product was not used for treatment.Visual evaluation of the returned sample noted one opened (without original packaging), used irrigation bulb syringe.Visual inspection of the sample noted that there was a strand of hair between the barrel and the bulb on the return sample.This does not meet the specification "no hair is permitted on the product.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The investigation is concluded, and no additional action is required at this time.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labelling could not have prevented the reported failure.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that there was hair found in the syringe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE IRRIGATION BULK / CLEAN, 50ML
Type of Device
IRRIGATION SYRINGE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12173456
MDR Text Key261701586
Report Number1018233-2021-04227
Device Sequence Number1
Product Code KYZ
UDI-Device Identifier10801741051378
UDI-Public(01)10801741051378
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number0935280
Device Catalogue Number0935280
Device Lot NumberNGEY0583
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/21/2021
Initial Date FDA Received07/15/2021
Supplement Dates Manufacturer Received01/20/2022
Supplement Dates FDA Received02/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-