(b)(4).The explanted device was returned and investigated.The implanted neurostimulator began experiencing errors shortly after stimulation was enabled.The investigation identified the issue as a corrupted measurement diagnostic which was interpreted by the software as an invalid value, resulting in no data uploaded to pdms after interrogation.The failure was confirmed to be a defect in a specific integrated circuit (asic) involved in the temporary storage of data prior to transmission (referred to as a storage buffer).The failure was not detected during manufacturing testing.Neuropace performs testing of the buffers with standard telemetry commands and random data, but testing is not able to cover every possible pattern.Neuropace is currently assessing the manufacturing process and vector test coverage.
|
The patient underwent initial placement of the rns system on (b)(6) 2021.On 5/3/2021, the treating center reported to neuropace that the patient had successfully interrogated their neurostimulator with the remote monitor and had uploaded the data to pdms (received a "successful transmission" message via the remote monitor).However no data was present in pdms.The patient was seen in the clinic on (b)(6) 2021 for troubleshooting and device programming (including re-installation of the neurostimulator firmware).The patient again attempted to interrogate on (b)(6) 2021, received the same message via the remote monitor, and again no data was seen on pdms.On (b)(6) 2021 the rns neurostimulator was replaced without complication.
|