Model Number VITD TOTAL |
Device Problems
Low Test Results (2458); Non Reproducible Results (4029)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/15/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer recalibrated the test and ran controls.Controls recovered within range.Patient samples 1-7 were then repeated.Refer to the attachment for the repeat values.The repeat values were in line with the clinical picture of the patients.
|
|
Event Description
|
The initial reporter stated they received discrepant results for eight patient samples tested with the elecsys vitamin d assay on a cobas e 411 immunoassay analyzer.No incorrect results were reported outside of the laboratory.Refer to the attachment for data from patient samples 1-7.All seven samples initially resulted in vitamin d values of < 3 ng/ml.To check if there were problems, the customer pulled a sample (sample 8) that had a validated result value of 43.89 ng/ml on (b)(6) 2021 and repeated it.The repeat result from this 8th sample was 6.46 ng/ml.The serial number of the e 411 analyzer is (b)(4).
|
|
Manufacturer Narrative
|
Medwatch field g4.Has been updated.
|
|
Manufacturer Narrative
|
Data was provided for 5 additional patient samples which had discrepant vitamin d results.The following data for these samples was provided: additional sample 1; run 1 = 8.87 ng/ml, run 2 = 56.9 n/ml.Additional sample 2; run 1 = < 3 ng/ml accompanied by a data flag, run 2 = 17.51 ng/ml.Additional sample 3; run 1 = 3.41 ng/ml, run 2 = 30.53 ng/ml.Additional sample 4; run 1 = < 3 ng/ml accompanied by a data flag, run 2 = 28.82 ng/ml.Additional sample 5; run 1 = 43.89 ng/ml, run 2 = 6.46 ng/ml, run 3 = 47.55 ng/ml.For this new patient data, no incorrect results were reported outside of the laboratory.
|
|
Manufacturer Narrative
|
For a calibration occurring on (b)(6)-2021, calibration signals for one calibrator level were within expected values.The signals for the second calibrator level were higher than expected.A new calibration performed on (b)(6)2021 with a different reagent pack had signals that were higher than expected.A new calibration performed on (b)(6)-2021 with the previous reagent pack had signals that were higher than expected.One control level was outside of range on (b)(6)-2021 and (b)(6)-2021.The second control level was within range, but borderline high on (b)(6)-2021.The second control level was within range on (b)(6)-2021 and (b)(6)-2021.The customer stored the reagent alternatively both on the analyzer and in the laboratory refrigerator.Upon review of the alarm trace, no relevant alarms were observed.The investigation could not identify a product problem.The cause of the event could not be determined.
|
|
Search Alerts/Recalls
|