MAUDE Adverse Event Report: DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Device Problems Overheating of Device (1437); Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/08/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Event Description
It was reported that the unit would power on via the evac but once the wash cycle was complete the unit would not power on via the wall.The event occurred during cleaning.There was some charring of the power cord plug that fits into the power entry module.No adverse event was reported as a result of this malfunction.
 
Event Description
There is no additional information.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that the power inlet module and cord showed signs of excessive heat.The power inlet module and cord were replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART WITH SMOKE EVACUATOR
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
MDR Report Key12189987
MDR Text Key262262843
Report Number0001954182-2021-00044
Device Sequence Number1
Product Code JCX
Combination Product (y/n)N
PMA/PMN Number
K172481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00514010200
Device Lot Number0025730
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/08/2021
Initial Date FDA Received07/19/2021
Supplement Dates Manufacturer Received08/30/2021
Supplement Dates FDA Received09/08/2021
Is This a Reprocessed and Reused Single-Use Device? No