Model Number N/A |
Device Problems
Overheating of Device (1437); Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that the unit would power on via the evac but once the wash cycle was complete the unit would not power on via the wall.The event occurred during cleaning.There was some charring of the power cord plug that fits into the power entry module.No adverse event was reported as a result of this malfunction.
|
|
Event Description
|
There is no additional information.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that the power inlet module and cord showed signs of excessive heat.The power inlet module and cord were replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|
|