The investigation of the pcb,front end,rohs was performed at getinge's national repair center(nrc) per procedure and to the ipc-a-610 standard with visual damage observed to connector j11, which has a broken pin.This board will not be tested due to the broken pin on j11.The board will be retained in the nrc per procedure.The investigation of 2nd board, pcb,front end, rohs was performed at getinge's nrc per procedure and to the ipc-a-610 standard with no visual damage observed.The nrc installed the board into the cardiosave test fixture and tested the board to factory specifications per procedure and the cardiosave service manual.The nrc could not verify the failure of the error code temp error on the screen.The board passed testing.The board will be sent to the supplier per procedure.
|