Catalog Number SGC0702
Device Problem Difficult or Delayed Positioning (1157)
Patient Problems Low Blood Pressure/ Hypotension (1914); Perforation of Vessels (2135)
Event Date 06/25/2021
Event Type  Injury  
Manufacturer Narrative
The customer reported the device is not returning.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report tissue damage, surgical intervention and prolonged hospitalization it was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4.It was noted enlarged atrium.There was resistance with the anatomy when the steerable guide catheter (sgc) was inserted through the inguinal region.Therefore, the sgc was expanded with a 20fr dilator and the sgc was re-inserted.The sgc was able to be advanced to the mitral valve, and the clip was deployed.After the sgc was removed without issues, the blood pressure dropped.A pressor agent was administered to stabilize the blood pressure.The patient remained hospitalized.One clip was implanted, reducing mr to 2.The next day, it was discovered that the iliac vein was damaged; and therefore, the vein was surgically treated.The patient is stable.There was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation was unable to determine a cause for the reported resistance with anatomy.The perforation appears to be related to the resistance with anatomy.The hypotension appears to be related to the perforation.Perforation and hypotension are listed in the instructions for use as known a possible complication associated with mitraclip procedures.The reported medication required, hospitalization and surgical intervention were results of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design, or labeling.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key12197128
MDR Text Key262469890
Report Number2024168-2021-06222
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/09/2022
Device Catalogue NumberSGC0702
Device Lot Number10210U122
Was Device Available for Evaluation? No
Date Manufacturer Received08/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other; Required Intervention;
Patient Age81 YR
Patient Weight50
-
-