Catalog Number 257589
Device Problem Contamination of Device Ingredient or Reagent (2901)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/24/2021
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Initial reporter phone #: (b)(6).Initial reporter fax #: (b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that prior to use bd bbl¿ 50 plates of schaedler k-v agar with 5% sheep blood they were discovered to be contaminated.The following information was provided by the initial reporter, translated from (b)(6) to english: the schaedler / schaed kv + 5% sb biplate agar plate was heavily contaminated and could not be used.
 
Event Description
It was reported that prior to use bd bbl¿ 50 plates of schaedler k-v agar with 5% sheep blood they were discovered to be contaminated.The following information was provided by the initial reporter, translated from german to english: the schaedler / schaed kv + 5% sb biplate agar plate was heavily contaminated and could not be used.
 
Manufacturer Narrative
H6: investigation summary: this statement is to summarize findings on the recent complaint 3137508 against schaedler agar / schaedler kv agar with 5% sheep blood (biplate), catalog number 257589, lot number 1104468.Event description: it was reported that 50 plates would have show contamination.Complaint history review: the complaint trends were reviewed for a period covering 12 months.For this product, several complaints were reported for contamination issues.Therefore, a trend was identified, and further investigation was performed.Batch history record (bhr) review: the batch history review was performed.As part of the investigation, the ingredients of this medium were reviewed.During this review, no deviation was detected.Sample analysis: picture sample was provided for evaluation.Based on the picture, the schaedler agar side was found to be contaminated.Retain samples were reviewed for this lot.During this review, contamination could be detected.Evaluation results: at this stage of our investigation, we have excluded any systemic failure in our manufacturing process.A material review board (mrb) was conducted which led to a situation analysis (sa) which eventually resulted in a recall.The performed investigation showed that the origin of the contamination was associated with an anaerobic microorganism growing at lower temperatures (<22°c).This microorganism was detected for the first time at the production site.Therefore, and as no contamination complaints were received in conjunction with the respective product within the past year this event was concluded to be a single incident.However, regarding the reported contamination, please note that this product does not have an sal (sterility assurance level) claim.It is filled aseptically, therefore an occurrence of a contamination event cannot be entirely ruled out.Investigation conclusion: based on the evaluation of the report the complaint is confirmed.An mrb was initiated and a sa was performed for contamination issue, which led to a recall.We would suggest that you set aside, and not use, any prepared plated media that does not meet the appearance specification as it is described on the bd certificate of analysis.This is consistent with industry recommendations for inspection of culture media prior to use (e.G.¿good practices for pharmaceutical microbiology laboratories¿, who technical report series, no.961, 2011, annex 2; chapter <1117> ¿microbiology best laboratory practices¿ the united states pharmacopeia; and the difco & bbl manual).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ SCHAEDLER K-V AGAR WITH 5% SHEEP BLOOD
Type of Device
CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL
Manufacturer (Section D)
BECTON DICKINSON GMBH
8-12 tullastrasse
heidelberg 69126
GM  69126
Manufacturer (Section G)
BECTON DICKINSON GMBH
8-12 tullastrasse
heidelberg 69126
GM   69126
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12197701
MDR Text Key262912774
Report Number9680577-2021-00032
Device Sequence Number1
Product Code JSJ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K760460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 12/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/15/2021
Device Catalogue Number257589
Device Lot Number1104468
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/10/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberIDS-21-4201
Patient Sequence Number1
-
-