(b)(4) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).No samples or photos were provided for evaluation.Unfortunately, as a result, bd was unable to verify the reported issue.The hair originates from an operator.The process has certain manual processes that may result in hair on the product.It is possible for associates to accidentally shed hair onto product as they are loading components into their corresponding containers/packages, although associates wear hair nets, gloves, safety glasses, and head covers, if worn improperly it could lead to the reported issue.The most probable root cause is inadequate gowning by associate(s) and/or preventive measures during the manufacturing of the product.Production record review was completed for batch/lot 1078273 and no non-conformance was noted during the manufacturing of this lot.An initiative is to replace lab coats with electrostatic lab coats to prevent hair from coming into the controlled manufacturing environment.No further action will be taken at this time and bd will continue to be tracked and trended.
|
It was reported by the facility representative that a hair was discovered inside the sealed package.Per email; i wanted to bring to your attention a situation regarding a compromised chloraprep clear, cat# lot#1078273.Before opening the package, a hair was discovered inside.Fortunately it was noted before opening.Can you please confirm if there have been any other incidents reported?.
|