• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY ASSURA DR ICD, US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY ASSURA DR ICD, US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD2357-40Q
Device Problems Pacing Asynchronously (1441); Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/28/2021
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer reference number: 2017865-2021-26250.Related manufacturer reference number: 2017865-2021-26253.It was reported in a routine clinical follow up that the implantable cardioverter defibrillator (icd) exhibited alerts for non-sustained ventricular oversensing, noise reversion, high pacing impedance, and pacemaker mediated tachycardia.Upon further inspection, the alerts were found to be indicative of right atrial (ra) and right ventricular (rv) lead issues.Both leads were dislodged and exhibited high pacing impedance, failure to sense, failure to obtain capture threshold, and prior episodes of noise oversensing.Twiddler¿s syndrome was suspected and was confirmed by the patient.The patient stated she felt well and had no symptoms.The patient did not schedule a follow up to address the issues.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12200166
MDR Text Key262609740
Report Number2017865-2021-26248
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734508094
UDI-Public05414734508094
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Model NumberCD2357-40Q
Device Catalogue NumberCD2357-40Q
Device Lot NumberA000106233
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/30/2021
Initial Date FDA Received07/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age51 YR
Patient Weight110
-
-