Model Number CSE-P-125-JA
Device Problems Crack (1135); Mechanical Problem (1384); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
On (b)(6) 2021, haemonetics manufacturing performed a visual inspection of the received bowl from the cell saver® elite set 125ml from (b)(6) and confirmed blood in the inner core with a crack in the inner core base.Visual inspection confirmed.The blood was seen inside the core in the bowl.Bowl decompression test was performed:.The blood outflow from the inner core and the outer core welding part was confirmed.Cause of the event is due to manufacturing.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due the event of the past report.
 
Event Description
On (b)(6) 2021, haemonetics was notified of a long hour returning error which was observed during a procedure in (b)(6), utilizing the cell saver® elite® autotransfusion system and cell saver® elite set - 125ml.There was no reported impact to patients' health.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER ELITE PLUS AUTOTRANSFUSION SYSTEM
Type of Device
CELL SAVER ELITE SET - 125ML, PRODUCT CODE: CAC
Manufacturer (Section D)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer (Section G)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer Contact
janice kiser
125 summer street
boston, MA 02110
MDR Report Key12201981
MDR Text Key263551374
Report Number1219343-2021-00114
Device Sequence Number1
Product Code CAC
UDI-Device Identifier10812747016537
UDI-Public(01)10812747016537(17)230820(10)0820066
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120586
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 06/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/24/2023
Device Model NumberCSE-P-125-JA
Device Lot Number0920058
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/21/2021
Initial Date FDA Received07/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-