Model Number CDVRA500Q
Device Problems Failure to Interrogate (1332); Inappropriate/Inadequate Shock/Stimulation (1574); Delayed Charge Time (2586)
Patient Problem Discomfort (2330)
Event Date 07/02/2021
Event Type  Injury  
Event Description
It was reported that the patient presented for initial implant of an implantable cardioverter defibrillator.During procedure, the physician charged the capacitor.Charging was slow.Upon charging, the device delivered a shock to the patient.The bluetooth telemetry was then shut off and the device was unable to be interrogated.Multiple programmers were used to interrogate the device but were unsuccessful.
 
Manufacturer Narrative
The reported event of communication anomaly was confirmed in the lab.Device testing showed the device was received below eol voltage.The cause of the field event is consistent with hybrid damage.The cause of the hybrid damage is undetermined.The cause of the inappropriate shock is undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12207669
MDR Text Key262866030
Report Number2017865-2021-26490
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public05415067031990
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Model NumberCDVRA500Q
Device Catalogue NumberCDVRA500Q
Device Lot NumberS000077866
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/04/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-