• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION AMIA AUTOMATED PD CYCLER SET; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION AMIA AUTOMATED PD CYCLER SET; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C5479
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/26/2021
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that three (3) amia cassettes had a small crack.This issue was identified during cassette testing on the cycler.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Correction: b4/f8: date of this report in the initial mdr is being corrected from blank to 07/22/2021.
 
Manufacturer Narrative
Additional information was added to h3 and h6.H10: the actual device was not available; however, a companion sample was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed on the companion sample including leak testing, clear passage testing, and clamp function testing, and no issues were observed.The reported condition was not verified with the companion sample.With the companion sample meeting specification and without the actual samples to analyze, the root cause of the event is unknown.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMIA AUTOMATED PD CYCLER SET
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key12211843
MDR Text Key262956532
Report Number1416980-2021-04502
Device Sequence Number1
Product Code FKX
Combination Product (y/n)Y
PMA/PMN Number
K151525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup,Followup
Report Date 08/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C5479
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/22/2021
Initial Date FDA Received07/22/2021
Supplement Dates Manufacturer Received07/22/2021
08/19/2021
Supplement Dates FDA Received07/22/2021
08/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
AMIA CYCLER; AMIA CYCLER
-
-