Additional information was added to h3 and h6.H10: the actual device was not available; however, a companion sample was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed on the companion sample including leak testing, clear passage testing, and clamp function testing, and no issues were observed.The reported condition was not verified with the companion sample.With the companion sample meeting specification and without the actual samples to analyze, the root cause of the event is unknown.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|