SMITH & NEPHEW, INC. GII SPC REAMTHRU CR TR SZ 5 LT; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71433345 |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/12/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that, a genesis ii spc reamthru cr tr size 5 left was found with a crack.As this was noticed upon field inspection, there was not patient involvement.
|
|
Manufacturer Narrative
|
H6: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned gii spc reamthru cr tr sz 5 lt confirmed the device is cracked on the back with multiple scratches in the metal.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|