ABBOTT GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDHFA500Q
Device Problems Inappropriate or Unexpected Reset (2959); Compatibility Problem (2960)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/08/2021
Event Type  malfunction  
Event Description
Following an mri, the device was found to be in back up mode.Technical support was contact and recommended reprogramming to resolve the event.The device was reprogrammed to resolve the event.The patient was stable.
 
Manufacturer Narrative
The reported event of back up mode following mri was confirmed.The device was not returned for analysis, however, abbott technical support did review session records.Based on the information received, the device event is consistent with the device entering power on reset (por) due to the mri settings not being enabled by the user prior to mri.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
MDR Report Key12217138
MDR Text Key263121654
Report Number2017865-2021-26602
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2022
Device Model NumberCDHFA500Q
Device Catalogue NumberCDHFA500Q
Device Lot NumberS000077210
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/08/2021
Initial Date FDA Received07/23/2021
Supplement Dates Manufacturer Received09/10/2021
Supplement Dates FDA Received09/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-