• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRISMATIK DENTALCRAFT, INC. SOFT NIGHT GUARD; THERMOFORM MOUTHGUARD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRISMATIK DENTALCRAFT, INC. SOFT NIGHT GUARD; THERMOFORM MOUTHGUARD Back to Search Results
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problem Hypersensitivity/Allergic reaction (1907)
Event Type  Injury  
Manufacturer Narrative
The device has not been returned.If/when the device is returned an investigation will be carried out and a supplemental report will be submitted.The exception of serial number as the device is manufactured by prescription.The device is manufactured by prescription and not implantable.
 
Event Description
It was reported that the patient experienced a reaction to the soft night guard.The type of allergic reaction noted was a burning sensation to the tongue.The reaction started on 6-23 and wore it half way through the night into 6-24.The duration of the event lasted 6 days, however the patient was still having discomfort.The provider advised to stop usage of appliance.The office was advised patient is "having a chemical burn" and seen by urgent care.
 
Manufacturer Narrative
The device has not been returned.However, the non-visual device investigation has been completed and the results are as follows: dhr results: the dhr was reviewed and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.Lot number was not provided - supplier (erkodent) could not review the associated material lot.Stock product reviewed results: no stock product was available for review since the device was fabricated per physician's prescription only.Investigation methods/results: customer stated the device has been returned but to date has not been received.However, the non-visual device investigation has been completed.Root cause: a root cause for this complaint cannot be explicitly determined.Ifu 9091 rev 3.0 (comfort h/s bite splint instruction for use) states to use only clear, cool water to wash the device.Ifu provides warning "do not clean or soak in mouthwash; do not use denture cleanser, hot water, alcohol, hydrogen peroxide; do not place in direct sunlight".It is possible that reactions could be caused by mouthwash, toothpaste, or soaking material.However, the customer did not provide the information regarding how the patient handled and maintained the device.Supplier erkodent reviewed the incident details and determined an allergic reaction would be possible if the device was built-up with acrylic.Glidewell research team and namsa conducted a series of testing on a similar thermoformed sleep device following iso 10993 (biological evaluation of medical devices) and the device was evaluated for potential cytotoxicity, skin irritation, delayed dermal contact sensitization and oral mucosal irritation.The test article was thermoformed with layers of erkodent material (erkoloc-pro and erkodur).The test results were listed below and summarized in biocompatibility report for sleep device (rpt 9733 rev 1.0).· for cytotoxicity testing, the test article extract showed no evidence of causing cell lysis or toxicity.· for skin irritation, there was no erythema and no edema observed on the skin of the animals treated with the test article.· for sensitization testing, the test article extracts showed no evidence of causing delayed dermal contact sensitization.· the test article showed nonirritant to the oral mucosa as compared to the control article.The device materials have been found to be biocompatible through the testing.There was no cytotoxic, sensitization, skin irritation, or oral mucosal irritation found in any of the test articles.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFT NIGHT GUARD
Type of Device
THERMOFORM MOUTHGUARD
Manufacturer (Section D)
PRISMATIK DENTALCRAFT, INC.
2144 michelson dr.
irvine CA 92612
Manufacturer (Section G)
PRISMATIK DENTALCRAFT, INC.
2144 michelson drive
irvine CA 92612
Manufacturer Contact
herbert crane
2144 michelson drive
irvine, CA 92612
9495021907
MDR Report Key12217507
MDR Text Key263155294
Report Number3011649314-2021-00250
Device Sequence Number1
Product Code MQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121365
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 01/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/28/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-