Model Number 4543
Device Problems Fracture (1260); Defective Device (2588); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2021
Event Type  Injury  
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, an evaluation of this lead was performed.The lead was returned in two segments, but the tip section was missing.Based on the finding of lead tip missing and tears in the insulation, this was likely explant damage.Based on the of the finding of fractured conductors likely mid-section of lead body.Fracture was a non specific fracture mode.Laboratory analysis confirmed the reported allegation.
 
Event Description
It was reported that this left ventricular (lv) lead had physical damage on the insulation and lead body.The lead also had a lead conductor fracture which was verified though x-ray and fluoroscopy.Subsequently, this lead was explanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
#12, rd. #698
dorado PR 00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12218428
MDR Text Key263171780
Report Number2124215-2021-20979
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526410994
UDI-Public00802526410994
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/08/2008
Device Model Number4543
Device Catalogue Number4543
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/14/2021
Initial Date FDA Received07/23/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/08/2006
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age81 YR
-
-