Model Number AS480
Device Problems Mechanical Problem (1384); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/25/2021
Event Type  malfunction  
Manufacturer Narrative
Root cause: alteration of staining in this case was due to improper operation of the liquid sensor board.The problem was solved by field service engineer with replacement of the part.Following the replacement, the instrument was fully operational within specifications, without errors and available for the user.Failure mode description: failure of the lsb, may result in a potential false negative result if aspiration position is intermittently too high or uneven staining occurs.This failure could be due to the vertical probe aspiration position being set above liquid level or not sufficiently deep into liquid, insufficient or no liquid aspirated, insufficient or no liquid dispensed, broken lsb, or liquid sensing system out of calibration.The lsb failure modes described above has the potential to alter staining.
 
Event Description
Customer complaint record reported the event as follows: some slides not being stained.No direct or indirect patient harm or user harm has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSTAINER LINK 48
Type of Device
AUTOSTAINER LINK 48
Manufacturer (Section D)
AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD.
no.1 yishun avenue 7
singapore north east, 76892 3
SN  768923
Manufacturer (Section G)
AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD.
no.1 yishun avenue 7
singapore north east, 76892 3
SN   768923
Manufacturer Contact
bryce portier
1834 state highway 71 west
cedar creek, TX 78612
4082185467
MDR Report Key12219735
MDR Text Key263398159
Report Number3003423869-2021-00247
Device Sequence Number1
Product Code KPA
UDI-Device Identifier05700572035497
UDI-Public05700572035497
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial
Report Date 07/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAS480
Device Catalogue NumberAS48030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/25/2021
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-