Model Number EG29-I10
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/24/2021
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: it was caused due to an excessive force applied on the insertion flexible tube (ift).In addition, we confirmed that bending rubber leak, the air/water nozzle looseness; however, these are not related to the alleged complaint.This report is being filed as part of the pentax backlog management plan.
 
Event Description
This event occurred at the time of before use.There was no report of patient harm.The ift is loosened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO GASTROSCOPE - I10 STANDARD
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key12222469
MDR Text Key263441713
Report Number9610877-2021-10365
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04961333248315
UDI-Public04961333248315
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K131902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 07/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG29-I10
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/24/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-