Model Number EEG-1200A |
Device Problems
Overheating of Device (1437); Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/24/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The monitor technician reported that there was a burning smell from the system.They did some troubleshooting and found that the power supply inside the computer was causing the smell.They replaced the part with the nihon kohden provided part number for the replacement power supply.Issue was resolved by the customer.This was not in patient use and no other devices being used with the system.
|
|
Event Description
|
The monitor technician reported that there was a burning smell from the system.They did some troubleshooting and found that the power supply inside the computer was causing the smell.They replaced the part with the nihon kohden provided part number for the replacement power supply.Issue was resolved by the customer.This was not in patient use.
|
|
Search Alerts/Recalls
|