Model Number TTNS3.00E4-1200 |
Device Problem
Flare or Flash (2942)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The product was returned and evaluated.Burnt traces were confirmed on the surface of the tip.The tip start time was 3:5:18 am on (b)(6) 2021 and was used for 905 treatments.The tip passed the flow test and failed the leak test.The tip failed visual inspection for burnt trace on the tip surface.Dielectric breakdown was observed.The tip passed the thermistor test.Functional testing was not performed due to the burnt trace on the tip surface.The plant evaluation is underway.
|
|
Event Description
|
A user facility reported they saw a light from the thermage tip and stopped using the tip.The tip was inspected and no abnormalities were noted.There was no patient injury.
|
|
Manufacturer Narrative
|
Service confirmed burnt radiofrequency trace on the tip membrane which most likely caused the reported issue during treatment.Investigation found glowing or sparking from the tip membrane are caused by stress concentrations on the flex assembly at the adhesive edge that damaged the radiofrequency trace, causing arcing and subsequent burn-through of the flex circuit membrane.Defects on the tip membrane can lead to a raise in temperature of the tip during treatment and can potentially cause patient burns.This evaluation confirms customer¿s account of possible light from the tip membrane during treatment.No patient injury was reported during treatment.A review of the manufacturing records showed all requirements were met.All treatment tips are visually inspected for defects during manufacturing and packaging.Final test verification specifications are acceptable.No non-conformities or anomalies found related to this complaint when reviewing the device history record for serial/lot number: (b)(6).
|
|
Search Alerts/Recalls
|