H4: the lot was manufactured from(b)(6) , 2020- (b)(6) , 2020.H10: the device was received containing fluid inside the bladder.Visual inspection was performed using the naked which showed no evidence of fluid flow coming out at the distal luer.Additional force prime was performed; however, flow was not observed.When the tubing line was cut to drain fluid, fluid was observed coming out from the tubing that was cut which indicated that an excess of solvent bonding was applied at the solvent-bonded junction between the tubing and distal luer.The reported condition was verified.The cause of the condition was due to excess solvent applied on the tubing during manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|