On (b)(6) 2021 the icd stopped transmitting to home monitoring.We tested all the possibilities and changed the cardio messenger of the patient.On (b)(6), follow up was not successful.Device in backup mode, unable to be reinitialized.Device was explanted and returned, but an explant date was not provided.
|
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.Next, the icd was subjected to an electrical analysis, revealing that the device could not be interrogated.Therefore, the device was opened and the inner assembly was inspected.The visual inspection of the inner assembly showed no anomalies.The measurement of the battery voltage showed a depleted battery.The current consumption of the electronic module was directly measured and found to be elevated.Further electrical investigations revealed that a capacitor of the electronic module was damaged, causing an increased current consumption, draining the battery.In conclusion, analysis revealed a damaged capacitor on the electronic module.There was no sign of any inconsistency during the manufacturing process which may be related to this observation.It is therefore assumed that the damage occurred after the device shipment, however, the date of occurrence was not determinable.
|