• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPINAL KINETICS LLC M6-C; ARTIFICIAL CERVICAL DISC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPINAL KINETICS LLC M6-C; ARTIFICIAL CERVICAL DISC Back to Search Results
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Unspecified Infection (1930)
Event Type  Injury  
Manufacturer Narrative
Additional information, including the serial number has been requested.Without a device, serial number or lot number, the device history records review could not be completed.This report is made by the manufacturer without prejudice and does not imply an admission of liability for the incident or its consequences.
 
Event Description
It was reported that and m6-c patient presented with fever one (1) week post op.Revision surgery performed, cultures taken, the device remains implanted.
 
Manufacturer Narrative
A1: pe 21-56.B4: 24-aug-2021.G3: 24-aug-2021.G6: follow-up #1.H2: additional information.H10: limited information was provided; notably, the device was not returned, no radiographs, and no lab reports were provided.It was not possible to assess the potential role of surgical technique or patient selection based on the paucity of information provided.No device, serial or lot number has been provided, therefore, a review of the lot history records for this device could not be performed.The risk management files were reviewed and no new risks were identified in the available reported information for this pe that require any changes to the current fmea, risk analysis, or labeling.This report is made by the manufacturer without prejudice and does not imply an admission of liability for the incident or its consequences.As such, we are reporting this incident to err on the side of caution.Spinal kinetics' post-market surveillance procedures require multiple attempts to gain as much detailed information as possible regarding the reported event.No further investigation can be performed.Based on the limited information provided, the device did not malfunction and it was not possible to determine the cause of the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M6-C
Type of Device
ARTIFICIAL CERVICAL DISC
Manufacturer (Section D)
SPINAL KINETICS LLC
501 mercury drive
sunnyvale CA 94085
MDR Report Key12243018
MDR Text Key264067823
Report Number3004987282-2021-00025
Device Sequence Number1
Product Code MJO
Combination Product (y/n)N
PMA/PMN Number
P170036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 08/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Date Manufacturer Received08/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-