• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION SUPERION INDIRECT DECOMPRESSION SYSTEM; PROSTHESIS, SPINOUS PROCESS SPACER/PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION SUPERION INDIRECT DECOMPRESSION SYSTEM; PROSTHESIS, SPINOUS PROCESS SPACER/PLATE Back to Search Results
Model Number 101-9812
Device Problems Difficult to Open or Close (2921); Material Integrity Problem (2978)
Patient Problem Insufficient Information (4580)
Event Date 03/29/2021
Event Type  malfunction  
Event Description
Device analysis performed on the returned superion indirect decompression spacer revealed that the spindle cap was completely separated from the implant body due to breaks found on the four weld joints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERION INDIRECT DECOMPRESSION SYSTEM
Type of Device
PROSTHESIS, SPINOUS PROCESS SPACER/PLATE
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
VERTIFLEX INC.
2714 loker ave. west
carlsbad CA 92010
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key12243766
MDR Text Key264058313
Report Number3006630150-2021-04196
Device Sequence Number1
Product Code NQO
UDI-Device Identifier00884662000543
UDI-Public00884662000543
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101-9812
Device Catalogue Number101-9812
Device Lot Number26375758
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/29/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-