MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number 37602 |
Device Problems
Failure to Deliver Energy (1211); Loss of Data (2903)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 03/26/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that stimulation settings disappeared non intentionally.The device was checked on (b)(6) 2021 and it was confirmed there were no problems.Telemetry was performed again on (b)(6) 2021, and a check was performed as usual, but it was confirmed that settings suddenly disappeared on this date.When checking on (b)(6) 2021, it was first identified that the stimulation settings were lost.Stimulation was not delivered between (b)(6) 2021 and (b)(6) 2021.The implantable neurostimulator (ins) was replaced on (b)(6) 2021 because the battery voltage was low and it was at the time of replacement.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received indicating the 8840 clinician programmer was used to interrogate the patient on 2021-(b)(6) , as well as on 2021-(b)(6).
|
|
Manufacturer Narrative
|
H3: analysis of the implantable neurostimulator (ins) (serial #: (b)(6) revealed no significant anomalies with the device where it experienced a normal end of life and telemetry and output were okay.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|